Specifications
Specifications
Imaging System
Laser Lifetime: 40,000 hours typical.
700 Channel Laser Source: Solid-state diode laser at 685 nm.
800 Channel Laser Source: Solid-state diode laser at 785 nm.
Detectors: Silicon avalanche photodiodes.
Scanning Speed: 5-40 cm/s.
Resolution: 21-337 µm.
Focus: Adjustable to be aligned to the top surface of the scan bed up to 4 mm above the scan bed to obtain the best signal-to-noise ratio of sample in membranes, gels, or microplates.
Dynamic Range:
4 logs (Manual); >6 logs (Auto).
Operating Specifications
Operating Conditions: For indoor use only; operating temperature and dew point not greater than 22 °C, non-condensing; altitude not to exceed 2000 m.
Environmental Conditions: Pollution Degree 2.
Power Requirements: Universal input range is between 100-240 VAC (voltage fluctuations not to exceed 10% of the nominal voltage); 4 Amp maximum; at 120V, 0.5 Amp typical when idle; 50/60 Hz.
Dimensions: 37 H × 53 W × 62 D cm (15 H × 21 W × 24.5 D inches). With the hood fully open, the Odyssey DLx requires 74 cm (29”) of vertical clearance.
Weight: 33 kg (72 lbs).
Network Protocol: TCP/IP.
Network Connection: Cat. 5e RJ45, 10BASE-T/100BASE-TX/1000BASE-T. Use only the supplied cable.
Detection Channel Summary
Channel | Excitation / Illumination | Emission Filter | Assay Category / Imaging Mode |
---|---|---|---|
800 | 785 nm | 812 – 832 nm | Membrane, Gel, In-Cell Western (includes On-Cell Western), Cell Analysis, , Custom |
700 | 685 nm | 710 – 730 nm | Membrane, Gel, In-Cell Western (includes On-Cell Western), Cell Analysis, , Custom |