Odyssey M Specifications

Imaging System Specifications

Unless otherwise specified, the following apply to Odyssey M Model 3350 and Odyssey M Model 3340.

Laser Source

The Odyssey M is a Class 1 laser product.

Laser SourceMaximum Laser OutputBeam Divergence
488 nm30 mW0.038°
520 nm50 mW0.046°
685 nm450 mW0.085°
785 nm1.2 W0.060°

Strong Electromagnetic Fields

Operating the Odyssey M in the presence of a strong electromagnetic field in the 100MHz to 400MHz frequency range has been shown to degrade the performance of the chemiluminescent imager. It is possible that some allowable faint bars or striping may be seen at slightly more than the background levels of the camera. When used in the expected environment, this interference should be uncommon.

LED

LED Risk Group: Exempt group in accordance with IEC 62471:2006. The LEDs do not pose any photobiological hazard.

Focus

Focus offset range: -1.00 mm to 5.00 mm

Pixel Resolution

Available imaging resolutions: 5 μm, 10 μm, 20 μm, 50 μm, 100 μm

Resolution for chemi images: 100 μm

Maximum Height of Samples

To image an item on the Odyssey M, the item must be 23 mm or less from the Scan Surface to the top of the item.

Operating Specifications

Operating Conditions

For indoor use only; operating temperature 15 - 30 °C and dew point not greater than 22 °C, non-condensing; altitude not to exceed 2000 m.

Transport Conditions

Shipping/transport conditions for the Odyssey M Imager should not exceed -40 °C to 65 °C.

Environmental Conditions

Pollution Degree 2

Power Requirements

  • Universal input range is between 100-240 VAC (voltage fluctuations not to exceed 10% of the nominal voltage); < 3 Amp maximum; At 120 V: 0.5 Amp typical when idle; 50-60 Hz.
  • Power Supply: One end of the provided power supply connects to a wall outlet and the other end connects to the imager. The power supply converts from wall outlet voltage to 24 VDC (5 Amp). Only use the provided power supply to plug in the imager. If you need another power supply, contact LICORbio.

Dimensions

  • 24" width × 30" depth × 15" height (61 cm width × 76 cm depth × 38 cm height)

  • Height with lid fully open: 28" (71 cm)

Scan Surface
  • Total Scan Area: 25 cm W × D 18 cm (9.8” W × 7.1” D)

  • Scan Area of Chemi Region (Odyssey M Model 3350 only): 15 cm W x 11 D cm (5.9” W × 4.3” D)

Dynamic Range

>6 logs for chemiluminescence (Model 3350) and fluorescence. This dynamic range is obtained in a single acquisition.

Weight

  • Odyssey M Model 3350: 121 lbs (55 kg)

  • Odyssey M Model 3340: 115 lbs (52 kg)

Network

  • Protocol: TCP/IP
  • Connection: Use the Category 6 network cable provided with the Odyssey M.

Detectors

  • sCMOS image sensor

  • Sensor for chemiluminescence (Model 3350): CCD

Expected Laser Lifetime

  • 685 nm: 20,000 hours

  • 785 nm: 20,000 hours

  • 488 nm: 40,000 hours

  • 520 nm: 40,000 hours

Light Sources

  • RGB LED (trans-illumination)

  • RGB LED (reflective illumination)

  • Solid-state diode laser at 488 nm

  • Solid-state diode laser at 520 nm

  • Solid-state diode laser at 685 nm

  • Solid-state diode laser at 785 nm

Odyssey M Detection Channel Summary

ChannelExcitation / IlluminationEmission FilterOdyssey M ChannelAssay Category / Imaging Mode
800785 nm816 - 840 nm785Ex-820EmMembrane, In-Cell Western (includes On-Cell Western), Cell Analysis, Microscope Slide, Custom
700685 nm721 - 740 nm685Ex-720EmMembrane, Gel, In-Cell Western (includes On-Cell Western), Cell Analysis, Microscope Slide, Custom
520520 nm570 - 610 nm520Ex-590EmMembrane, Gel, In-Cell Western (includes On-Cell Western), Cell Analysis, Microscope Slide, Custom
520A520 nm721 - 740 nm520Ex-720EmCustom
520B520 nm816 - 840 nm520Ex-820EmCustom
488488 nm519 - 543 nm488Ex-530EmMembrane, Gel, In-Cell Western, Cell Analysis, Microscope Slide, Custom
488A488 nm570 - 610 nm488Ex-590EmGel, Custom
488B488 nm721 - 740 nm488Ex-720EmCustom
488C488 nm816 - 840 nm488Ex-820EmCustom
ChemiNANAChemiMembrane, Custom (luminescent assays)
RGB Epi470 525 630 EpiNARGBEpiCustom
630 Epi631 - 659 nmClear Window630EpiMembrane
525 Epi500 - 530 nmClear Window525EpiMembrane
470 Epi447 - 472 nmClear Window470EpiCustom
RGB Trans470 525 630 TransNARGBTransMicroscope Slide, Custom
630 Trans631 - 659 nmClear Window630TransGel, Absorbance Assay
525 Trans500 - 530 nmClear Window525TransCustom
470 Trans447 - 472 nmClear Window470TransGel, Absorbance Assay