Odyssey XF Imager Specifications

Specifications

Operating Conditions: For Indoor use only; operating temperature 15 - 35 °C and dew point < 22 °C, non-condensing; maximum operating temperature may be reduced at elevations above 2000 m.

Environmental Conditions: Pollution Degree 2.

Dynamic Range: 22 bit (>6 logs)

Depth of Field for Best Sample Focus: 6 mm

Patented FieldBriteTM Technology: CV <3%

600 Channel Light Source: Light source at 520 nm

700 Channel Laser Source: Solid-state Laser Diode at 685 nm

800 Channel Laser Source: Solid-state Laser Diode at 785 nm

Laser Lifetime: 20,000 hours

Detector Type: Low-noise CCD. Thermoelectrically cooled.

Acquisition Times:

600, 700 and 800 nm channels: 30 s, 2 min, 10 min plus advanced variable time feature.

Chemiluminescence channel: 30 s, 2 min, 10 min, 60 min plus advanced variable time feature.

Focusing: None needed.

Sensitivity: Similar to Odyssey® DLx Imaging System

Wavelength Maxima for Excitation/Emission Bands:

Channel

Excitation

Emission

600 nm

520 nm

600 nm

700 nm

685 nm

730 nm

800 nm

785 nm

830 nm

Chemiluminescence

none

Visible

Dyes: Optimized for LI‑COR® IRDye® 680RD and IRDye® 800CW reagents, and chemiluminescent substrates.

Image Display Options: Pseudocolor, positive and negative grayscale, single color (red, green or blue), or two colors with overlapping fluorescence displayed in a third color.

Capacity: One imaging tray containing one blot or gel.

Field of View: 10 cm D × 12 cm W (3.9” D × 4.7” W) at the surface of the imaging bed.

Power Requirements: Universal input range is between 100-127 VAC and 200-240 VAC (voltage fluctuations not to exceed 10% of the nominal voltage); 50/60 Hz; at 120V, 4 Amp maximum; at 120V, 0.3 Amp typical when idle. Overvoltage Category II.

Computer Connection Cable: Cat. 5e RJ45, 10BASE-T/100BASE-TX/1000BASE-T. Use only the supplied cable.

Size: 67.3 H × 41.4 W × 47 D cm (26.5 H × 16.3 W × 18.5 D inches). Depth with imaging drawer open is 59.7 cm (23.5”).

Weight: 27 kg (60 lbs).

The Odyssey XF Imager is certified as a Class 1 laser product. Hazardous laser radiation is not emitted outside the instrument. Radiation emitted inside the Odyssey XF Imager is confined within protective housings and external covers. A series of safety interlocks ensures that laser beams cannot escape during any phase of user operation.

Detection Channel Summary

ChannelExcitation / IlluminationEmission FilterAssay Category / Imaging Mode
800785 nm816 - 846 nmMembrane, Gel
700685 nm716 - 746 nmMembrane, Gel
ChemiNANAMembrane
600520 nm573 - 637 nmMembrane, Gel