Odyssey EMSA Kit

Components

TubeReagentVolumeStorage
110X Binding Buffer, 100 mM Tris, 500 mM KCl, 10 mM DTT; pH 7.5 500 μL-20 °C
225 mM DTT, 2.5% Tween® 20500 μL-20 °C
3Poly (dI•dC), 1 μg/μL in 10 mM Tris, 1 mM EDTA; pH 7.5125 μL-20 °C
4Sheared Salmon Sperm DNA, 0.5 μg/μL in 10 mM Tris, 1 mM EDTA; pH 7.5125 μL-20 °C
550% Glycerol500 μL-20 °C
61% NP-40500 μL-20 °C
71 M KCl500 μL-20 °C
8100 mM MgCl2500 μL-20 °C
9200 mM EDTA, pH 8.0500 μL-20 °C
1010X Orange Loading Dye500 μL-20 °C

Storage

Items are shipped on dry ice. Upon receipt, items should be stored as indicated in Components.

Description

Gel shift assays, also called electrophoretic mobility shift assays (EMSA), provide a simple method to study DNA-protein interactions. This assay is based on the principle that a DNA-protein complex will have different mobility during electrophoresis than non-bound DNA. These shifts can be visualized on a native acrylamide gel using labeled DNA to form the DNA-protein binding complex. To date, protocols require labeling DNA by (1) radioisotope, (2) digoxygenin, or (3) biotin. The Odyssey Infrared EMSA Kit, coupled with IRDye® 700 EMSA oligonucleotides, is an excellent alternative method to radioisotopic and chemiluminescent detection methods for EMSA analysis and visualization (4,5). Using IRDye EMSA reagents, assays can be completed in less than two hours with no gel transfer or film exposure. The gel doesn’t need to be removed from the glass plates for imaging.

Protocol Timeline

EMSA protocol timeline from beginning to the completion of the protocol, which takes approximately 1.5 hours
Figure 43. This workflow graphic shows the infrared EMSA timeline with IRDye 700 EMSA oligonucleotides.

Additional Required Materials

  1. IRDye® end-labeled DNA. IRDye 700-labeled, double-stranded DNA oligonucleotides are available from LI-COR for various transcription factors. See www.licor.com/EMSAoligos for the most up-to-date selection, or see IRDye Labeling of DNA Fragments in Section V. for further details on how to obtain IRDye end-labeled oligonucleotides.

    DNA must be end-labeled. Internal labels (as generated by random priming) will inhibit DNA-Protein complex formation.

  2. Ultrapure, nuclease-free water

  3. Polyacrylamide gel (5% gel is a good starting point) in Tris-acetate, Tris-borate, or Tris-glycine-EDTA and corresponding electrophoresis buffer

  4. Electrophoresis unit

  5. Odyssey DLx, Odyssey CLx, or Odyssey M Imager

 

General Methodology for Infrared Electrophoretic Mobility Shift Assay (IR-EMSA)

IRDye® Labeling of DNA Fragments

IRDye 700-labeled, double-stranded DNA oligonucleotides are available from LI-COR for various transcription factors (see www.licor.com/EMSAoligos for the most up-to-date selection). To obtain IRDye end-labeled DNA fragments for other transcription factors of interest, IRDye infrared dye-labeled oligos can be used. It is critical that the DNA fragment is end-labeled rather than having dye incorporated into the DNA. Dye incorporation into the DNA fragment interferes with formation of the DNA-Protein complex.

Synthetic oligonucleotides 5’ end-labeled with IRDye 700 or IRDye 800 are available from several suppliers. Oligonucleotides are manufactured in single-strand form; therefore, both sense and antisense DNA oligonucleotides must be purchased.

Both oligonucleotide sequences should be end-labeled with the same IRDye. There is a significant decline (~70%) in signal intensity when using only one end-labeled oligonucleotide.

Once oligonucleotides are obtained, they need to be annealed to form a double-stranded DNA fragment. Oligonucleotides are annealed by placing the oligonucleotide set in a 100 °C heat block for 5 minutes. Leave the oligonucleotides in the heat block and turn it off to slowly cool to room temperature.

Binding Reaction

A universal binding condition that applies to every protein-DNA interaction is not recommended, since binding conditions are specific for each protein-DNA interaction. You should establish binding reaction conditions for each protein-DNA pair. The Odyssey Infrared EMSA Kit provides all the reagents necessary to optimize an IR-EMSA binding reaction.

Optimized binding reactions for all infrared dye-labeled EMSA-ready oligonucleotides are provided in their pack inserts, along with a suggested positive control nuclear extract. All other IR-EMSA binding reactions will need further optimization. Optimization can be done by adding supplementary components to the binding buffer supplied in this kit. This can be done using the binding reaction setup in Table 1. Reactions should be set up and added in the order listed in Table 1.

In some cases, DNA control reactions (no protein) were observed to have lower signal than reactions containing protein. This may be due to lower stability of the dye in certain buffer conditions. The addition of DTT and Tween® 20 to all reactions reduces this phenomenon.

Reactions are incubated to allow protein binding to DNA. Time required for binding is the same as when radioactively-labeled DNA fragments are used. A typical incubation condition is 20 - 30 minutes at room temperature. Because IRDye infrared dyes are sensitive to light, it is best to keep binding reactions in the dark during incubation periods (e.g., put tubes into a drawer or simply cover the tube rack with aluminum foil). After the incubation period, 2 μL of 10X Orange Loading Dye (Tube 10) should be added to the 20 μL EMSA reaction.

Do not use blue loading dye (e.g., bromophenol blue), because it will be visible when imaged on an Odyssey Imager.

Table 1This table shows a sample IR-EMSA binding reaction optimization experiment.
ComponentReaction
 12345678
Ultra Pure water14 μL13 μL 12 μL12 μL12 μL 12 μL12 μL12 μL
10X Binding Buffer (Tube 1)2 μL2 μL2 μL2 μL2 μL2 μL2 μL2 μL
25 mM DTT/2.5% Tween® 20 (Tube 2)2 μL2 μL2 μL2 μL2 μL2 μL2 μL2 μL
1 μg/μL Poly (dI•dC) (Tube 3)1 μL1 μL1 μL1 μL1 μL1 μL1 μL1 μL
Optional components        
0.5 μg/μL Salmon Sperm DNA (Tube 4)--1 μL-----
50% Glycerol (Tube 5)---1 μL----
1% NP-40 (Tube 6)----1 μL---
1 M KCl (Tube 7)-----1 μL--
100 mM MgCl2 (Tube 8)------1 μL-
200 mM EDTA (Tube 9)-------1 μL
IRDye end-labeled oligo (50 nM)1 μL1 μL1 μL1 μL1 μL1 μL1 μL1 μL
Protein Extract (5 μg/μL)-1 μL1 μL1 μL1 μL1 μL1 μL1 μL
Total Volume20 µL20 µL20 µL20 µL20 µL20 µL20 µL20 µL

Electrophoresis

  • For best results, perform electrophoresis in dark (simply put a cardboard box over the electrophoresis apparatus).

  • Separation of protein-DNA complexes is usually performed by loading binding reactions onto native polyacrylamide gels. Percentage of the gel depends on the protein size and DNA fragment, but a 5% gel is a good starting point. Electrophoresis is usually performed at 10 V/cm at room temperature or at 4 °C in Tris-acetate, Tris-borate, or Tris-glycine-EDTA gel and buffer.

Imaging

Find imaging information for your Odyssey Imager and imaging software at licor.com/support.

Quantification

  • When quantifying Odyssey-generated EMSA data, great care must be taken in experimental design and analysis.

  • One of the benefits of using Odyssey for EMSA analysis is that it provides an easy method for quantification. However, there are issues to consider when using Odyssey to quantify EMSA results. The primary issue is that the free DNA fragment has much less signal than the DNA fragment when bound to a protein, making quantification of the unbound DNA inaccurate. The addition of DTT/Tween® 20 to the binding reaction stabilizes the dye and reduces this phenomenon.

  • In addition, it is unrealistic to perform quantification analyses under the assumption that the free DNA band in the control containing DNA only (no extract), should equal the sum of the signals of the free and bound DNA in the samples where the protein-DNA binding reaction occurs. Using end-labeled oligonucleotide duplexes as the DNA source and nuclear extract as a protein source render this assumption impractical, due to the non-specific binding that occurs from using a nuclear extract. Oligonucleotides can also complicate quantification because the free oligonucleotides form a smear rather than a tight band. This makes it more difficult to assign an intensity value to the bands.

Mobility Shift Example Experiment Using IRDye® 700 AP-1 Consensus Oligonucleotide

Prepare and Pre-Run Gel

  1. Use a pre-cast 5% TBE native acrylamide gel, or prepare a native acrylamide gel in 0.5X TBE.

  2. Pre-electrophorese gel in 0.5X TBE for 30 minutes at 70 V (for a 9 x 7 cm gel) or 10 V per cm gel.

  3. While gel is pre-running, proceed to Binding Reaction steps.

Binding Reaction

  1. Reactions should be set up and added in the order listed.

    ComponentReaction
     1234
    Ultra Pure water12 μL12 μL9 μL9 μL
    10X Binding Buffer (Tube 1) 2 μL 2 μL 2 μL 2 μL
    25 mM DTT/2.5% Tween® 20 (Tube 2)2 μL2 μL2 μL2 μL
    1 μg/μL Poly (dI•dC) (Tube 3)1 μL1 μL1 μL1 μL
    1% NP-40 (Tube 6)1 μL1 μL1 μL1 μL
    100 mM MgCl2 (Tube 8)1 μL1 μL1 μL1 μL
    200X Consensus Competitor Oligo0 μL0 μL1 μL0 μL
    200X Mutant Competitor Oligo 0 μL 0 μL 0 μL1 μL
    IRDye 700 AP-1 Consensus Oligonucleotide 1 μL1 μL1 μL1 μL
    HeLa 4-hr Serum Response Nuclear Extract (2.5 μg/μL)-2 μL2 μL2 μL
    Total Volume20 µL20 µL20 µL20 µL
  2. Gently mix tube by tapping the sides of the tubes.

  3. Briefly centrifuge sample into the bottom of the tube.

  4. Place tubes in box or drawer protected from light, and incubate for 20 minutes at room temperature.

  5. Add 2 μL of 10X Orange Loading Dye (Tube 10) to each reaction tube. Gently pipette to mix.

Electrophoresis

  1. Turn off pre-running gel current.

  2. Flush gel wells with 50 μL of electrophoresis buffer.

  3. Load 20 μL of each sample reaction into the wells of the gel.

  4. Turn on current to 70 V (or 10 V per cm of gel)

  5. Cover gel box with dark box.

  6. Electrophorese for 1 hour or until orange dye migrates to the bottom of the gel.

Example Data

Figure 44. IRDye 700 AP-1 EMSA.

Troubleshooting

Weak or No Signal

Possible CauseSolution / Prevention
Did not add DTT/ Tween® 20 to binding reactionAdd 1 μL of 25 mM DTT/ 2.5% Tween® 20 to binding reaction
Not enough IRDye® dye- labeled DNA usedIncrease amount of IRDye dye-labeled DNA added to the reaction
Target DNA degradedVerify integrity of DNA
Imaged in the wrong channel of the Odyssey When using IRDye 700-labeled DNA, turn on the 700 nm laser

No Shifted Bands Detected or Weak Signal

Possible CauseSolution / Prevention
Incorrect Focus OffsetAdjust the acquisition Focus Offset to equal the thickness of the glass plate plus half the thickness of the gel and scan again
DNA/Protein complex disrupted due to heat or vortexing

Run gel with cooled buffer

Do not vortex binding reaction

Not enough extractAdd more extract to reaction
Degraded extract

Minimize freeze-thaw cycles

Use protease inhibitors

System not fully optimizedUse additives in the kit to determine their effects on binding efficiency

Spots or Speckling

Possible CauseSolution / Prevention
Contamination on glass surfacesClean glass gel plates and the Odyssey scanning surface with isopropanol

References

  1. Wolf, S.S., Hopley, J.G., and Schweizer, M. (1994) The application of 33P-labeling in the Electrophoretic mobility shift assay. Biotechniques 16, 590 - 592.

  2. Suske, G., Gross, B., and Beato, M. (1989) Non-radioactive method to visualize specific DNA-protein interactions in the band shift assay. Nucleic Acids Res. 17, 4405.

  3. Ludwig, L.B., Hughes, B.J., and Schwartz, S.A. (1995) Biotinylated probes in the electrophoretic mobility shift assay to examine specific dsDNA, ssDNA or RNA-protein interactions. Nucleic Acids Res. 23, 3792 - 3793.

  4. Li, Y., Ahmed, F., Ali, S., Philip, P.A., Kucuk, O., and Sarkar, F.H. (2005) Inactivation of nuclear factor κB by soy isoflavone genistein contributes to increased apoptosis induced by chemotherapeutic agents in human cancer cells. Cancer Res. 65: 6934 - 6942.

  5. Geddie, M.L., O’Loughlin, T.L., Woods, K.K., and Matsumura, I. (2005) Rational design of p53, and intrinsically unstructured protein, for the fabrication of novel molecular sensors. J. Biol. Chem. (M508149200).